Description
Model
G-CO2
Electrical requirements
110V to 220V, 50/60Hz 1-Phase
Cooling
Lowest temperature: -70°C Cooling type: direct cooling Refrigerant: CO2 Injection cycle during operation: 2 seconds, at 15 seconds interval
Thermostat
PT100
Cryo valve
Imported from Germany
Backup battery
48 hours
Safety
High temperature alarm Low battery alarm
Connection to freezer
Thru one inch port on ultra-low freezer
Maintenance
Test backup system once per month manually to prevent failure